| Products |
|
|
Utilizing advanced fluorescence technology, the algaewatch On-line Fluorometer offers real-time algal biomass data with unmatched accuracy and sensitivity.
The algaewatch is the ideal tool for water resource managers and scientists interested in real-time monitoring of water quality. Easy to use and install, the algaewatch provides a continual indication of the algal biomass in water, thus allowing for the early detection of algal blooms, detection of nuisance species, and a measure of daily and seasonal fluctuations in the algal community.
Algaewatch is an on-line monitor that detects algal biomass through chlorophyll fluorescence. Chlorophyll a, the primary photosynthetic pigment in all plants, is a fluorescent molecule. Where chlorophyll-containing organisms are small enough (as with phytoplankton or microalgae), fluorescence may be measured directly (in vivo chlorophyll detection) without extraction or chemical treatment. | |
- Internal Data Logger: 20,000-point capacity with programmable sampling interval and Windows compatible software.
-
- Control Outputs: Two programmable AC, one 0–5V DC output.
- Easy Maintenance Design: Clean-out port and brushes to allow simple routine maintenance.
-
- Optional Calibration Kit: Includes calibration standard and accessories for easy calibration and performance verification.
| |
- Detect the onset of algal blooms
-
- Eliminate or decrease frequency of cell counting
-
- Collect data on algal community dynamics
-
- Make more effective use of treatment chemicals
-
- Detect the presence of taste and odor causing species
-
- Improve filter run times
| |
| Sensitivity |
0.1 µg/L chlorophyll a |
| Linear Range |
0–200 µg/L chlorophyll a |
| Accuracy |
±0.2 µg/L or 3% of the reading,
whichever is greater |
| Repeatability |
±3% |
| Calibration |
1 point calibration (standard and blank) |
| Power |
90–250 VAC/200–250 VAC, 50/60 Hz, 5 amps |
| Relays |
2 x 90-250 VAC/200-250 VAC, 50/60 Hz, 5 amps
(fused at 3.15 amps ea. Type “F”), 1 x 0-5V DC |
| Signal Output |
One 4–20 mA, isolated |
| Dimensions |
8"W x 4"D x 15"H; 20cm x 10cm x 38cm |
| Weight |
5 lbs; 2.3 kg |
| Enclosure |
Approximates NEMA 4X |
| Maximum Water Pressure |
100 psi |
| Ambient Temperature |
40°F to 120°F; 4°C to 49°C |
| Maximum Sample Temperature |
140°F; 60°C |
| Relative Humidity |
0–100% |
|
|  |
| |
|
|