Hydrolab provides the most accurate Blue-Green Algae sensor available on a multiprobe. Available in two forms, one for detecting phycocyanin (fresh water), and one for detecting phycoerythrin (marine water) | |
- Available in two forms, one for detecting phycocyanin (fresh water), and one for detecting phycoerythrin (marine water)
- Ultra-compact size designed specifically for integration into the Hydrolab DS5X, DS5, and MS5
- Available with solid Secondary Standards to provide a quick and simple method to verify the sensor's stability over time
- The Secondary Standard can be adjusted to correlate to a known Blue-Green Algae concentration
- Three auto-selected gain ranges provide a wide measurement range of 100 to 2,000,000 cells/mL for either phycocyanin or phycoerythrin
| |
- Real-time measurement:
- Identifies potential algal blooms before they become problematic, allowing time for corrective action
- Less expensive and more timely than cell counting or visual inspection
- Plus, like all other fluorometers available on a Hydrolab sonde:
- Provides the most accurate measurement of phycocyanin or phycoerythrin because of electronic filtration of ambient light, efficient optical coupling and quality optical components
- Incredibly fast response time through electronic filtration of ambient light
- Excellent turbidity rejection due to small sample volume design and high quality optical filters
- Cost-optimized for affordability and value
| |
Optical Characteristics:
| Light Source: |
Light Emitting Diode |
| Detector: |
Photodiode |
| Excitation Wavelength: |
Phycocyanin 590 nm
Phycoerythrin 525 nm |
| Emission Wavelength: |
Phycocyanin 650 nm
Phycoerythrin 570 nm |
Product Data:
Specs
Minimum Detection Limit:
100 cells/mL
Dynamic Range:
Low sensitivity: 150-2,000,000 cells/mL
Med. sensitivity: 150-200,000 cells/mL
High sensitivity: 150-20,000 cells/mL
Accuracy:
+/- 3% for signal level equivalents of 1 ppb rhodamine WT dye or higher using a rhodamine sensor
Resolution:
20 cells/mL
Sensor housing:
Stainless steel:
Standard housing for typical fresh water applications.
Titanium option:
Corrosion-resistant housing for use in aggressive saline environments such as oceans, bays and estuaries.
|
|  |