The Enviro-T, easily integrated with data collection systems, provides a 4 - 20 mA signal output proportional to the relative concentration of the fluorophore of interest in the sample water. The Enviro-T is a rugged, 24/7 sampling device that provides maximum performance, minimal maintenance, and solid-state reliability.
|
Excitation: 488 nm (rated at 20,000hr life), 640 nm(rated at 20,000hr life)
Laser Profile: 15 x 75 microns
Scatter Detection: Forward (0 degrees, +/-15), Side (90 degrees, +/-15)
Emission Detection: 4 colors, user swappable optical filters
Standard set installed:
FL1 530/30nm
FL2 585/40nm
FL3 >670nm
FL4 660LP
Optical Alignment: Fixed alignment, no maintenance required
Sensitivity Adjustment: Fixed, no adjustment required
Flow Cell: 200 micron ID quartz capillary
Minimum Particle Size: 0.5μm
Minimum Sample Size: 50μl (in 1.5mL conical tube)
Nominal Flow Rate -variable: 10 to 100 μl/min
Fixed Flow Rates & Core Sizes:
Slow -(Flow = 14μL/min ; Core = 10 μm)
Medium -(Flow = 35μL/min ; Core = 16 μm)
Fast -(Flow = 66μL/min ; Core = 22 μm)
Recommended Sheath Fluid: 0.2 μm filtered DI water
Core Diameter: User selectable from 5 to 40 microns
Fluorescence Sensitivity; MESF*: FITC <150, PE <100
Fluorescence Linearity: 2+/-0.05% for chick erythrocyte nuclei (CEN)
Fluorescence Precision: <3% CV for CEN
Absolute Counts: Yes
Speed: 10,000 events/second maximum
Signal Processing: 24 bit A/D conversion
Computer Interface: USB 2.0
Warm-up Time: Less than 5 minutes
*MESF values determined using Thermo Scientific Cyto-Cal™ Multifluor Plus Violet Beads.
Power: 100-240 VAC, 50/60 Hz
Typical Power Consumption: 154 VA
Heat Output: 240 Btu/hour maximum output
Operating Ranges: 15-30ºC; <80% relative humidity
Instrument Size: 11”H x 14.75”W x 16.5”D (27.9 x 37.4 x 41.9 cm)
with Fluid Tanks: 11”H x 21.5”W x 16.5”D (27.9 x 54.6 x 41.9 cm)
Weight: 30 lbs (13.6 kg)
Fluid Tank Capacity: 2L Sheath Fluid, 2L Waste, 250 mL Cleaner, 250 mL Decontamination |